Ankom 2000 Series Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Ankom 2000 Series. ANKOM 2000 Series User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Operator’s Manual

Operator’s Manual 2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986-8090 Fax: (315) 986-8091 www.ankom.com Rev E 1/14/15

Page 2

Operator’s Manual pg. 10 Rev E 1/14/15 Safety Precautions Hazardous Pressure – Do NOT open the Vessel Lid during operation.

Page 3 - Table of Contents

Operator’s Manual Rev E 1/14/15 pg. 11 Instrument Installation Site Requirements To install and operate the ANKOM2000 Fiber Analyzer

Page 4

Operator’s Manual pg. 12 Rev E 1/14/15 6. If you are using Cubetainers for your chemicals, install the Cubetainer Shelf shown below (Opti

Page 5 - Operating Environment

Operator’s Manual Rev E 1/14/15 pg. 13 Fiber Analysis Support Items The following support items are needed to perform the fiber analy

Page 6 - Contact Information

Operator’s Manual pg. 14 Rev E 1/14/15 This page intentionally left blank

Page 7 - Instrument Description

Operator’s Manual Rev E 1/14/15 pg. 15 ADF Analysis ADF Calculation ADF contained within a food or feed sample can be calculated usin

Page 8

Operator’s Manual pg. 16 Rev E 1/14/15 5. Place an empty filter bag in the Bag Holder in an open position. 6. Tare the weight of the e

Page 9

Operator’s Manual Rev E 1/14/15 pg. 17 ADF Analysis Procedure using the ANKOM2000 Fiber Analyzer To perform ADF analysis on prepared

Page 10 - Safety Precautions

Operator’s Manual pg. 18 Rev E 1/14/15 8. Press ENTER on the Keypad and follow the prompts on the Display to set up the instrument for AD

Page 11 - Instrument Installation

Operator’s Manual Rev E 1/14/15 pg. 19 NDF Analysis NDF Calculation NDF contained within a food or feed sample can be calculated usin

Page 12 - IMPORTANT

This page intentionally left blank

Page 13 - Fiber Analyzer

Operator’s Manual pg. 20 Rev E 1/14/15 5. Place an empty filter bag in the Bag Holder in an open position. 6. Tare the weight of the e

Page 14

Operator’s Manual Rev E 1/14/15 pg. 21 NDF Analysis Procedure using the ANKOM2000 Fiber Analyzer To perform NDF analysis on prepared

Page 15 - ADF Analysis

Operator’s Manual pg. 22 Rev E 1/14/15 11. Press the arrow keys on the Keypad until you see “Select NDF” on the Display. Select NDF  v

Page 16

Operator’s Manual Rev E 1/14/15 pg. 23 IMPORTANT Do NOT place bags in the oven until the acetone has completely evaporated. 20. Pl

Page 17

Operator’s Manual pg. 24 Rev E 1/14/15 This page intentionally left blank

Page 18

Operator’s Manual Rev E 1/14/15 pg. 25 Crude Fiber Analysis Crude Fiber Calculation Crude Fiber contained within a food or feed sampl

Page 19 - NDF Analysis

Operator’s Manual pg. 26 Rev E 1/14/15 5. Place an empty filter bag in the Bag Holder in an open position. 6. Tare the weight of the e

Page 20

Operator’s Manual Rev E 1/14/15 pg. 27 Crude Fiber Analysis Procedure using the ANKOM2000 Fiber Analyzer To perform Crude Fiber analy

Page 21

Operator’s Manual pg. 28 Rev E 1/14/15 IMPORTANT The Crude Fiber analysis will use the following settings:  40 minute Acid Digestion 

Page 22

Operator’s Manual Rev E 1/14/15 pg. 29 16. Remove the samples from the oven and place them in a Desiccant Pouch. IMPORTANT Do NOT

Page 23

Table of Contents Introduction ...

Page 24

Operator’s Manual pg. 30 Rev E 1/14/15 This page intentionally left blank

Page 25 - Crude Fiber Analysis

Operator’s Manual Rev E 1/14/15 pg. 31 Flush Procedure The Flush procedure allows you to clean the system with water. This procedure

Page 26

Operator’s Manual pg. 32 Rev E 1/14/15 This page intentionally left blank

Page 27

Operator’s Manual Rev E 1/14/15 pg. 33 Periodic Maintenance Initial Maintenance (after 10 hours of operation) After the first 10 ho

Page 28

Operator’s Manual pg. 34 Rev E 1/14/15 If you see a leak If you see a leak, follow steps 6 – 9 in the Initial Maintenance procedure above.

Page 29

Operator’s Manual Rev E 1/14/15 pg. 35 Check Agitation System and Bag Suspender The agitation system and bag suspender should be chec

Page 30

Operator’s Manual pg. 36 Rev E 1/14/15 Check Bag Suspender The bag suspender should be checked every three to six months or if

Page 31 - Flush Procedure

Operator’s Manual Rev E 1/14/15 pg. 37 Appendix A – ADF Method Acid Detergent Fiber in Feeds - Filter Bag Technique (for A2000 and A2

Page 32

Operator’s Manual pg. 38 Rev E 1/14/15 Calculations % ADF (as-received basis) = 100 x (W3 – (W1 x C1)) W2 Where: W1 = Bag tare weight

Page 33 - Periodic Maintenance

Operator’s Manual Rev E 1/14/15 pg. 39 Appendix B – NDF Method Neutral Detergent Fiber in Feeds - Filter Bag Technique (for A2000 and

Page 34

This page intentionally left blank

Page 35

Operator’s Manual pg. 40 Rev E 1/14/15 Calculations % NDF (as-received basis) = 100 x (W3 – (W1 x C1)) W2 Where: W1 = Bag tare weight

Page 36 - Troubleshooting

Operator’s Manual Rev E 1/14/15 pg. 41 Appendix C – Crude Fiber Method Crude Fiber Analysis in Feeds - Filter Bag Technique (for A200

Page 37 - Appendix A – ADF Method

Operator’s Manual pg. 42 Rev E 1/14/15 Precision Results of the collaborative study (see Tables 1&2) indicate the precision (S

Page 38

Operator’s Manual Rev E 1/14/15 pg. 43 Collaborative Laboratory No.Rep Whole CornCattle FeedAlfalfaWhole SoyPoultry StarterCalf

Page 39 - Appendix B – NDF Method

Operator’s Manual pg. 44 Rev E 1/14/15 This page intentionally left blank

Page 40

Operator’s Manual Rev E 1/14/15 pg. 45 Appendix D – Parts & Assemblies 101.2 Bag Holder X45 Desiccant Pouch 693 Bag Suspender

Page 41

Operator’s Manual pg. 46 Rev E 1/14/15 A07 Water Filter Assembly A21 Water Valve Assembly A22 Vent Valve Assembly A19 B or A Valve Asse

Page 42

Operator’s Manual Rev E 1/14/15 pg. 47 963 Power Supply 5653 On/Off Switch 687 Fuse 15A 6035 Fuse 10A 6151 SICK Fiber Optic L

Page 43 - Rev E 1/14/15 pg. 43

Operator’s Manual pg. 48 Rev E 1/14/15 Appendix E – Wiring Diagrams (1 of 4)

Page 44

Operator’s Manual Rev E 1/14/15 pg. 49 Appendix E – Wiring Diagrams (2 of 4)

Page 45

Operator’s Manual Rev E 1/14/15 pg. 5 Introduction ANKOM Technology designs, manufactures, and markets instruments and suppor

Page 46

Operator’s Manual pg. 50 Rev E 1/14/15 Appendix E – Wiring Diagrams (3 of 4)

Page 47

Operator’s Manual Rev E 1/14/15 pg. 51 Appendix E – Wiring Diagrams (4 of 4)

Page 48

Automation saves time and money! ANKOM Technology is an international company with products that include… TDF Dietary Fiber Analyzer  Autom

Page 49

Operator’s Manual pg. 6 Rev E 1/14/15 Contact Information At ANKOM Technology we are committed to your total satisfaction and therefore alw

Page 50

Operator’s Manual Rev E 1/14/15 pg. 7 Instrument Description General Description The ANKOM2000 Fiber Analyzer is designed to effici

Page 51

Operator’s Manual pg. 8 Rev E 1/14/15 Power Switch Port B – Used for Crude Fiber Base solution, AD solution, and Diluted Amylase Po

Page 52 - Fax: (315) 986-8091

Operator’s Manual Rev E 1/14/15 pg. 9 Internal Components A2000 Computer (part # A09) Check Valve Assembly Pressure Transducer (p

Commentaires sur ces manuels