Operator’s Manual
Rev L 11/12/14
2052 O’Neil Rd, Macedon NY 14502
Telephone: (315) 986-8090
Fax: (315) 986-8091
www.ankom.com
Vue de la page 0
1 2 3 4 5 6 ... 27 28

Résumé du contenu

Page 1 - Operator’s Manual

Operator’s Manual Rev L 11/12/14 2052 O’Neil Rd, Macedon NY 14502 Telephone: (315) 986-8090 Fax: (315) 986-8091 www.ankom.com

Page 2

Operator’s Manual pg. 10 Rev L 11/12/14 4. Attach the HCl Zinc Filter Assembly (part# H36) to the “Vent Hose” as shown below. Position the h

Page 3 - Table of Contents

Operator’s Manual Rev L 11/12/14 pg. 11 Instrument Setup This step is extremely important to fill the waste tank in the instrument with w

Page 4

Operator’s Manual pg. 12 Rev L 11/12/14 This page intentionally left blank

Page 5 - Operating Environment

Operator’s Manual Rev L 11/12/14 pg. 13 Hydrolysis Procedure Total Fat Calculation The hydrolysis procedure is done prior to fat extractio

Page 6 - Contact Information

Operator’s Manual pg. 14 Rev L 11/12/14 Sample Categorization For best results, it is important to determine the correct ratio of Diatomaceous

Page 7 - Instrument Description

Operator’s Manual Rev L 11/12/14 pg. 15 Sample Categorization Data from ANKOM Lab (sorted by DE Weight) Sample Type Fat % DE Weight Sample

Page 8 - Safety Precautions

Operator’s Manual pg. 16 Rev L 11/12/14 Sample Preparation Procedure To avoid fat migration from one filter bag to another during t

Page 9

Operator’s Manual Rev L 11/12/14 pg. 17 Moist Samples a. Add 1/2 of the recommended amount of DE to the filter bag. b. Using a spatula,

Page 10

Operator’s Manual pg. 18 Rev L 11/12/14 11. Include two blank bags (C1 and C2) filled with approximately 0.5g – 0.75g of DE. Blank bags serv

Page 11

Operator’s Manual Rev L 11/12/14 pg. 19 Hydrolysis step-by-step procedure using the ANKOMHCl Hydrolysis System To perform hydrolysis on pr

Page 12

This page intentionally left blank

Page 13 - Hydrolysis Procedure

Operator’s Manual pg. 20 Rev L 11/12/14 16. Press START on the Keypad. The ANKOMHCl Hydrolysis System will automatically heat and maintain th

Page 14

Operator’s Manual Rev L 11/12/14 pg. 21 Periodic Maintenance Cleaning the exterior of the instrument At the end of each day use an ammonia

Page 15

Operator’s Manual pg. 22 Rev L 11/12/14 5. Press the tab and lift the Fuse Holder up. 6. Replace the fuses and close the Fuse Holder compl

Page 16

Operator’s Manual Rev L 11/12/14 pg. 23 Troubleshooting & Replacement Parts The ANKOM Technology web site has the most current trouble

Page 17

Operator’s Manual pg. 24 Rev L 11/12/14 This page intentionally left blank

Page 18

Operator’s Manual Rev L 11/12/14 pg. 25 Appendix A – Electrical Diagram (pg. 1 of 3)

Page 19

Operator’s Manual pg. 26 Rev L 11/12/14 Appendix A – Electrical Diagram (pg. 2 of 3)

Page 20

Operator’s Manual Rev L 11/12/14 pg. 27 Appendix A – Electrical Diagram (pg. 3 of 3)

Page 21 - Periodic Maintenance

Automation saves time and money! ANKOM Technology is an international company with products that include… TDF Dietary Fiber Analyzer  Automates

Page 22

Table of Contents Introduction ...

Page 23

This page intentionally left blank

Page 24

Operator’s Manual Rev L 11/12/14 pg. 5 Introduction ANKOM Technology designs, manufactures, and markets instruments and support p

Page 25

Operator’s Manual pg. 6 Rev L 11/12/14 Contact Information ANKOM Technology is committed to your total satisfaction and therefore always avail

Page 26

Operator’s Manual Rev L 11/12/14 pg. 7 Instrument Description General Description The ANKOMHCl Hydrolysis System is designed to release

Page 27

Operator’s Manual pg. 8 Rev L 11/12/14 Safety Precautions Hot Surfaces – Do not touch the Vessel Lid during operation. The surface can e

Page 28

Operator’s Manual Rev L 11/12/14 pg. 9 Instrument Installation and Setup Site Requirements To install and operate the ANKOMHCl Hydrolysis